8-4
4. Select
The Intervals screen closes.
5. Schedule levels for each time and day, as necessary:
a. Use the arrow keys to navigate to a place in the grid indicating the day of the week
b. Select one or more buttons for the level or levels to schedule.

Setting up Required QC Ranges

Use this screen to enter the following information for the controls you use for Required
QC analysis:
Control name
Lot number
Expiration date
Target ranges for each level
Required QC ranges support both an active and trial lot. Only the active lot is used to
satisfy the Required QC schedule. The trial lot is used to evaluate new materials.
To define the schedule your system uses for automatically analyzing controls, refer to
Setting up AutomaticQC Schedule‚ page 8-5.
1. At the Setup screen, select
NOTE:
an active lot for a level and scan in new lot information for the same level, it becomes
the trial lot.
When you scan the barcode, the system makes the button corresponding to that level
available. If a level button is available, that barcode is already scanned.
2. Scan the
When you scan the barcode for a new lot of controls, the system displays a message
prompting you to keep the existing target ranges.
3. Scan the
chart.
This displays the Range Entry screen. This screen displays the control name, level, lot
number, and expiration date under the banner portion of the screen. It also displays the
parameters and their ranges and allows you to edit the ranges. Refer to Editing
Required QC Ranges‚ page 8-5.
4. At the Range Entry screen, scan the barcode for each parameter to enter the target
ranges for this level.
5. Select
02087462 Rev. V
.
Save
and time period for the level or levels.
A check mark on the button means that you scheduled that level. Select the button
again to remove the check mark.
When you scan a lot for a level, it becomes the active lot. If you already have
catalog number barcode
lot number barcode
.
Save
Rapidlab 1200 Operator's Guide: System Configuration
QC > Required QC Ranges
on the QC Expected Values chart.