Rapidlab 1200 Operator's Guide: Troubleshooting

Removing Obstructions from the CO-ox Sample Path

Use the following procedure to remove obstructions from the CO-ox sample path
where you see fluid restrictions and obstructions, or suspect problems based on
D code messages.
BIOHAZARD:
Refer to Appendix A, Protecting Yourself from Biohazards for recommended
precautions when working with biohazardous materials.
CAUTION:
without first discharging static buildup. Touch the inner surface of the
measurement module frame to discharge static buildup.
1. Remove the reagent cartridge and the wash cartridge:
a. Select
b. Remove both the cartridges.
c. Leave the doors open.
2. If the obstruction is between the sample chamber and the measurement
module connector, replace the CO-ox sample tubing:
a. Disconnect the sample tubing from the connector on top of the sample
1
2
3
Figure 6-15
Wear personal protective equipment. Use universal precautions.
Do not remove or return the sensors to the measurement module
Status > Diagnostics > Cartridges > Eject Both Cartridges
chamber.
Sample chamber connector
Fluid detector
Measurement module connector
CO-ox Sample Path
6-31
.
02087462 Rev. V