Rapidlab 1200 Operator's Guide: Maintenance

Relocating the System

Use this procedure to move the system for use in another area.
WARNING:
damage to the system, remove power from the system as described in this
procedure.
CAUTION:
cartridges are installed. Cartridges installed in the system remain stable for 6
hours without power.
WARNING:
mixture with air, O
explosive environment. Refer to Protecting Yourself from Electrical Hazards‚
page A-3 for recommended precautions when working around electricity.
1. Select
2. When the system displays the Shutdown screen, turn the power off and
disconnect the power cord from the electrical outlet.
3. Disinfect the exterior surfaces of the system.
Refer to Cleaning and Disinfecting the Exterior Surfaces‚ page 5-2.
4. Move the system to its new location.
5. To restore power to the system, connect the power cord to the electrical outlet.
6. Turn the power on.
Do not proceed while the power is on. To prevent electrical shock or
Do not remove power from the system for more than 6 hours if
Do not operate the system in the presence of flammable anesthetic
, or nitrous oxide. The risk of explosion exists in a potentially
2
.
Status > Shutdown
5-49
02087462 Rev. V