Rapidlab 1200 Operator's Guide: System Specifications
Precision on manual aqueous quality control materials was estimated using a
minimum of 3 Rapidlab 1245 systems. Controls were measured over a minimum
of 10 days.
Precision on the aqueous quality control materials in the AutomaticQC module
was estimated using a minimum of 3 Rapidlab 1245 systems over a minimum of
10 days. AutomaticQC is automatically scheduled to run (single sample)
independently 3 times per day.
The following tables summarize the results of the Rapidlab 1245 system precision
for manual QC and AutomaticQC:
Table E-16
Parameter
pH
pH
pH
pCO
2
pCO
2
pCO
2
pO
2
pO
2
pO
2
tHb
tHb
tHb
FO
Hb
2
FO
Hb
2
FO
Hb
2
FCOHb
FCOHb
FCOHb
FMetHb
FMetHb
FMetHb
FHHb
FHHb
FHHb
*
WRSD = within-run standard deviation
Rapidlab 1245 Manual QC Precision Results
Level
n
Mean
1
161
7.119
2
161
7.328
3
160
7.529
1
161
72.6
2
162
40.7
3
160
22.9
1
161
147.3
2
162
97.7
3
160
24.8
1
132
17.8
2
132
14.2
3
132
7.8
1
132
80.0
2
132
93.3
3
132
60.1
1
132
1.0
2
132
1.8
3
132
16.4
1
132
15.9
2
132
2.0
3
132
7.1
1
132
3.2
2
132
2.9
3
132
16.4
*
WRSD
Total SD
0.001
0.002
0.001
0.002
0.001
0.003
0.42
1.14
0.29
0.56
0.21
0.36
0.69
1.85
0.85
1.56
0.73
1.07
0.05
0.14
0.05
0.11
0.03
0.06
0.15
0.21
0.14
0.29
0.22
0.32
0.14
0.19
0.11
0.22
0.18
0.26
0.11
0.15
0.10
0.21
0.18
0.41
0.05
0.08
0.05
0.11
0.08
0.14
02087462 Rev.V