Siemens Rapidlab 1200 Operator's Manual page 364

Table of Contents

Advertisement

E-4
Table E-5
Parameter
ctO
(Hb)
2
The following table lists the units, reporting ranges, and display resolutions for the
temperature corrected and respiratory parameters calculated by the Rapidlab 1200
systems.
Table E-6
Parameter
pH(T)
+
H
(T)
pCO
(T)
2
pO
(T)
2
pO
(A-a)(T)
2
pO
(a/A)(T)
2
RI(T)
sp/ t(T)
sp/ t(est, T)
The following table lists the units, reporting ranges, and display resolutions for the
metabolic parameters calculated by the Rapidlab 1200 systems.
Table E-7
Parameter
-
HCO
3
-
HCO
3
ctCO
2
BE(B)
BE(ecf)
02087462 Rev.V
CO-ox Module Parameters
Units
mL/dL
mL/L
mmol/L
Temperature Corrected and Respiratory Parameters
Units
pH
nmol/L
mmHg
kPa
mmHg
kPa
mmHg
kPa
%
decimal fraction
%
decimal fraction
%
decimal fraction
%
decimal fraction
Metabolic Parameters
Units
mmol/L
act
mmol/L
std
mmol/L
mmol/L
mmol/L
Rapidlab 1200 Operator's Guide: System Specifications
Reporting Range
0–40.0
0–400
0–17.9
Reporting Range
6.000–8.000
10.0–1000.0
5.0–250.0
0.67–33.33
0.0–800.0
0.00–106.66
0–999.9
0.00–133.31
0–100
0.00–1.00
0–2000
0.00–20.00
0–100
0.00–1.00
0–100
0.00–1.00
Reporting Range
0.0–99.9
0.0–99.9
0.0–99.9
-99.9–99.9
-99.9–99.9
Resolution
0.1
1
0.1
Resolution
0.001
0.1
0.1
0.01
0.1
0.01
0.1
0.01
1
0.01
1
0.01
1
0.01
1
0.01
Resolution
0.1
0.1
0.1
0.1
0.1