6-48

D2 Codes

D2 Excessive Drift: pO
Problem
Sensor drift is
beyond predefined
limits during
calibration
operations.
02087462 Rev. V
Rapidlab 1200 Operator's Guide: Troubleshooting
pCO
2
2
Corrective Action
Correct any reagent problems such as D23 codes before you
attempt to correct this D code.
If the sensor was newly installed or if maintenance was
recently performed, allow sufficient time for the sensor to
equilibrate.
Perform additional calibrations.
If a D2 for pCO
, with a drift greater than 4.0 mmHg,
2
occurs twice in a 4-hour period and no drift occurs for pO
perform QC analysis.
If the pCO
result is out of range, replace the sensor.
2
Deproteinize the sample path.
Refer to Deproteinizing the Sample Path‚ page 5-4.
Check that the sensors are installed in the correct location
and are aligned, the O-rings are in place, and the
spring-loaded latch is closed.
If necessary, remove and install the sensors correctly.
If the problem persists, replace the sensor.
If the problem persists, call for technical assistance.
,
2