D23 Reagent Flow Error: 6

Problem
The system detects
an air leak in the
fluidic path during
a calibration.

D23 Reagent Flow Error: 7, 8

Problem
Wash fluid was not
detected during
wash after a
calibration or
sample analysis.
7 calibration
8 sample analysis.
02087462 Rev. V
Rapidlab 1200 Operator's Guide: Troubleshooting
Corrective Action
1.
Perform the highlighted test(s) in
.
Functions
2.
Perform a leak test.
Refer to Performing the Leak Test‚ page 6-37. If the test
fails, check the tubing for leaks, and that the tubing is
correctly installed in the pinch valve, the O-rings are in
place, and the spring-loaded latch is closed. For Rapidlab
1245 and 1265 systems, verify that the CO-ox sample
chamber is installed correctly.
3.
If the problem persists, replace the reagent cartridge.
Perform the highlighted test(s) in
.
Functions
4.
If the problem persists, replace the wash cartridge.
Perform the highlighted test(s) in
.
Functions
5.
If the problem persists, call for technical assistance.
Corrective Action
1.
Perform the highlighted test(s) in
.
Functions
2.
Perform a leak test.
Refer to Performing the Leak Test‚ page 6-37. If the test
fails, check the tubing for leaks, and that the tubing is
correctly installed in the pinch valve, the O-rings are in
place, and the spring-loaded latch is closed. For Rapidlab
1245 and 1265 systems, verify that the CO-ox sample
chamber is installed correctly.
3.
If the problem persists, replace the wash cartridge.
Perform the highlighted test(s) in
.
Functions
4.
If the problem persists, call for technical assistance.
Diagnostics > Fluidic
Diagnostics > Fluidic
Diagnostics > Fluidic
Diagnostics > Fluidic
Diagnostics > Fluidic