Rapidlab 1200 Operator's Guide: Quality Control
2. Introduce the QC sample:
a. Scan the barcode on the QC ampule, if necessary.
b. Invert the QC ampule several times and then open the ampule.
c. Prepare the sample.
If the sample device is...
an ampule
a syringe
d. Insert the sample device into the sample port.
e. Select
f.
3. If prompted, enter your operator ID and then select
4. View the results on the screen and on the printed reports.
If the Auto Print option is turned on in Setup, the system automatically prints
the QC sample report.
5. When you finish viewing the results, select

Accessing QC Results

When the system completes QC analysis, the system displays the Results screen.
For a Required QC analysis and for Unscheduled QC analysis when you scan the
barcode on the ampule, the Results screen immediately under the banner contains
the following information:
QC type
Level
Lot
For AutomaticQC samples, the system displays the level.
.
Analyze
When prompted, remove the sample device from the sample port and then
select
.
Continue
Then...
attach a Quick adapter to the ampule.
draw the QC sample into the syringe.
Continue
Continue
4-5
.
.
02087462 Rev. V