Final procedures
Service PC and measuring instruments
Saving the configuration file
NOTE
Main menu:
1.
2.
Reset the exposure counter
Main menu:
1.
2.
Deleting the error buffer
Main menu:
1.
2.
Saving the MAMMOMAT Parameters
Main menu:
1.
2.
NOTE
Remove the measuring instruments
1.
2.
3.
Checking the protective grounding resistance
Using the protective ground wire tester, measure the resistance between the protective
ground terminal in the stand (Mains input terminal) and all metallic parts of the equipment
which the patient/operator may come in contact with, incl. all object tables.
The protective ground resistance must not exceed 0.1 Ω.
Siemens AG
Medical Solutions
Make sure that the backup floppy is inserted.
Select Configuration
Press <F2> to save data.
Select Service
Reset exposure counter.
Delete the error memory with <Y>.
Select Service
Delete error buffer.
Delete the error memory with <Y>.
Select Backup
Copy installation area to floppy
Leave the program with <F10> when the backup is ready.
Keep the backup floppy with other documentation for the
MAMMOMAT Novation
Turn OFF the MAMMOMAT.
Remove the service PC.
Remove connected measuring instruments.
Save config file.