Cleaning the Sample Path

Use this procedure to clean the sample path with bleach.
You need the following materials for this task:
10% bleach solution
syringe
lint-free tissue or swabs
deproteinizer
test/blank reference sensor (TB5)
test/blank Glucose sensor (TB4)
test/blank Lactate sensor (TB4)
BIOHAZARD:
Refer to Appendix A, Protecting Yourself from Biohazards for recommended
precautions when working with biohazardous materials.
CAUTION:
sensor with the test/blank reference sensor (TB5) before cleaning.
CAUTION:
conditioner, or deproteinizer. Replace the biosensors with the test/blank sensors
(TB4) before conditioning the sensors or deproteinizing the sample path. Reinstall
the biosensors within 2 hours.
CAUTION:
sensors.
CAUTION:
without first discharging static buildup. Touch the inner surface of the
measurement module frame to discharge static buildup.
1. Select
2. To start the procedure, select
3. Replace the reference sensor and the biosensors with the test/blank sensors:
a. Lift the door to the measurement module.
02087462 Rev. V
Wear personal protective equipment. Use universal precautions.
Do not expose the reference sensor to bleach. Replace the reference
Do not expose the glucose and lactate biosensors to bleach,
Do not use alcohol to perform this procedure. Alcohol can damage the
Do not remove or return the sensors to the measurement module
Status > Maintenance > Deproteinizing the Sample Path
Refer to Preparing for Maintenance Procedures‚ page 5-1.
Rapidlab 1200 Operator's Guide: Maintenance
.