Troubleshooting The Purple Parameter Error; Troubleshooting The Ampule Qc - Siemens Rapidlab 1200 Operator's Manual

Table of Contents

Advertisement

Rapidlab 1200 Operator's Guide: Troubleshooting

Troubleshooting the Purple Parameter Error

Error
The parameter
is purple with a
line through it
and QC! is
displayed in
the lower left
corner.

Troubleshooting the Ampule QC

Use the following recommendations to help resolve issues with failed QC
performed from an ampule:
Perform only one QC analysis for each ampule.
Performance claims are based on one QC analysis per ampule.
Always use the Quick adapter when analyzing QC samples.
Analyze the QC material immediately after opening the ampule.
Prolonged contact with the atmosphere affects pH, pCO
Ensure that the correct target ranges are defined in Setup.
Store QC material at a constant temperature.
Do not handle ampules excessively.
Variation in results may occur due to changes in temperature.
Do not use expired QC materials.
Do not analyze QC materials that contain bubbles.
Use the correct level of QC material that is scheduled for analysis.
Probable
Cause
Corrective Action
Analyze all the currently scheduled controls.
Required QC
analysis was
Refer to Performing Required QC Sample
Analysis‚ page 4-1.
not performed
when
If the scheduled controls do not contain the
scheduled.
necessary levels, analyze QC for the
appropriate levels.
Refer to Performing Unscheduled QC Sample
Analysis‚ page 4-4.
If the results are within range, turn the
parameter on.
Refer to Restoring Parameters‚ page 4-12.
6-3
, and pO
results.
2
2
02087462 Rev. V