Rapidlab 1200 Operator's Guide: System Overview and Intended Use

Reagent Module

The reagent module holds the reagents and creates the fluid path for samples,
calibrators, and wash fluid.
The reagent module consists of the following components:

reagent cartridge

cartridge interface frame
reagent manifold
reagent door
Reagent Cartridge
The reagent cartridge contains the following 2 calibrators, each in foil bags:
Calibrator Volume
Slope
200
You use these calibrators, as well as the RCx and wash in the wash cartridge, to
calibrate the system. The electrolytes, pH, glucose, lactate, and gases in the
reagents are NIST traceable.
The following table lists the targeted calibration points for each analyte in the
reagents:
Analyte
pH
pCO
2
pO
2
+
Na
+
K
++
Ca
-
Cl
Glu
Lac
tHb
*
fixed point via electronic zero
Ingredients
160 mL
gases (oxygen, carbon dioxide, nitrogen), salts (alkali
halides), organic buffers, catalyst, and surfactant
460 mL
gases (oxygen, carbon dioxide, nitrogen), salts (alkali
halides), organic buffers, glucose, lactate, surfactant, and
preservative
Cal Point
6.8
35 mmHg
154 mmHg
116 mmol/L
4.0 mmol/L
1.25 mmol/L
98 mmol/L
180 mg/dL
2 mmol/L
0 g/dL
Slope Point
7.4
70 mmHg
*
0 mmHg
159 mmol/L
8.0 mmol/L
0.62 mmol/L
69 mmol/L
0 mg/dL
0 mmol/L
15 g/dL
02087462 Rev. V
1-9