Using The Calibration Report To Identify Problems - Siemens Rapidlab 1200 Operator's Manual

Table of Contents

Advertisement

Rapidlab 1200 Operator's Guide: Troubleshooting
Check the solution level in the affected sensor and remove bubbles.
The K
should be completely full without any bubbles.
Check the sensor contacts and measurement block for salt buildup.
Check behind the sensors for fluid leaks that can come from the reference
sensor.
Check that the sensors are installed in the correct order and are aligned, the
O-rings are in place, and the spring-loaded latch is closed.
If the pH or Na
Check for cracks or leaks in the sample tubing, and replace the tubing if
necessary.
Replace the affected sensor, if necessary.

Using the Calibration Report to Identify Problems

The full calibration report shows the measurement and drift values. Rapidlab 1200
systems use a 2-point calibration curve. The two points on the curve are the Cal
Point and Slope Point.
The values in the Meas column represent the actual measured value of the
current calibration.
The values in the Drift column represent the difference between the measured
value from the current calibration and the value assigned to the analyte during
the manufacture of the cartridge.
If the system cannot report a value during a calibration, the result is blank on
the report and the D code and explanation prints at the bottom of the report
The following symbols can display with the results on the printed report.
Table 6-1
This symbol...
↑ or ↓
+
++
, Ca
, and Cl¯ sensors should be nearly full. The pH and Na
+
sensor have a D2 or D3, condition the sensors.
Symbols ↑ or ↓
Indicates that...
the result is above or below the drift limit.
6-5
+
sensors
02087462 Rev. V