List of Error codes

A full list of error codes is given below::
CODE
Description
Af
EXPOSURE TERMINATED BY FILM AEC
Ms
EXPOSURE TERMINATED BY mAs INTEGRATOR
CODE
Fault description
Rp
EARLY PUSH BUTTON RELEASE
Mi
LOW mAs SELECTION WITH AEC
Bt
EXPOSURE TERMINATED BY BACK UP TIMER
T1
ANODE STARTER FAILED DURING PREPARATION
T2
FILAMENT FAILED DURING PREPARATION
R1
ANODE STARTER FAILED DURING EXPOSURE
R2
FILAMENT FAILED DURING EXPOSURE
Operating or system errors with sample pulse x-ray emission only
CODE
Fault description
A0
BREAST TOO DENSE
A1
AEC DETECTOR OVER EXPOSED
CODE
Fault description
D0
DOOR OPEN
C0
ABSENCE OF COLLIMATOR CONE
Nc
CASSETTE NOT INSERTED
Ci
CASSETTE IN AT POWER ON
Ce
CASSETTE ALREADY EXPOSED
T0
NOT ACTIVE
CODE
Fault description
Gs
GAS SPRING DEFECTIVE
Pb
CHECK BRAKE PUSH BUTTONS
Pa
CHECK ARM PUSH BUTTON
Pc
CHECK FOOT PEDAL SWITCH
Fi
FILTER POSITION ERROR
Me
MIRROR POSITION ERROR
Pi
BRAKE POSITION SWITCHES MW5 , MW6 FAILURE
Mt
BRAKE MOTOR MT3 FAILURE
HU
TUBE TERMAL LIMIT REACHED
FUSES and blocks ( multiple fault are shown as code addition)
CODE
Fault description
80
INVERTER CONTROL BOARD
40
FILAMENT POWER SUPPLY
20
ROTATING ANODE STARTER
10
INVERTER POWER
08
F4 PCB 92-082
04
F2 + 15P PCB 92-082
02
F2 + 15° PCB 92-082
01
ANODE NOT ROTATING
MAMMOMAT Balance
Troubleshooting Guide
Exposure terminated without problems.
Exposure terminated with problems.
Operating errors without x-ray emission
System errors and or faults
SPB7-115.840.01
Page 22 of 62
Rev.02
10.04
Troubleshooting