b. Push a clot-removal line through the preheater from right to left, opposite to the
1
2
Figure 6-10
CAUTION:
damaging the gas sensors, do not exert excessive pressure to the gas sensor
membranes.
4. Inspect the pO
sensors for obstructions:
a. Remove the gas sensors one at a time and inspect each one for obstructions.
02087462 Rev. V
direction of sample flow, until you see the clot-removal line at the sample
connector.
Preheater
Sample connector
Rapidlab 1200 Systems Sample Path
Do not push the clot-removal line through the gas sensors. To avoid
and pCO
2
2
Rapidlab 1200 Operator's Guide: Troubleshooting
(gas), sample ground/temperature, pH, and the reference