Siemens Rapidlab 1200 Operator's Manual page 36

Table of Contents

Advertisement

1-16
The sample enters the measurement module at the sample connector and moves through
the preheater, fluid detector 1, and into the sensors for measurement.
1
2
3
4
Figure 1-19
The preheater warms the sample to 37°C and the measurement block ensures a constant
temperature of 37°C. The sample connector provides the fluidic path for the sample from
the sample entry components in the reagent cartridge to the measurement module.
The sensors detect analytes present in the sample and form the sample path. The system
moves the sample through the sensors and the fluid detector 2, into the pinch valve tubing,
and through the pressure detector bubbler.
02087462 Rev. V
Rapidlab 1200 Operator's Guide: System Overview and Intended Use
Preheater
Fluid detector 1
Sample connector
Sensors
Measurement Module