6-14
d. Carefully pull the reagent manifold out slightly, then straight down, to avoid the
1
2
3
4
5
Figure 6-1
5. Place the reagent manifold on a stable, flat surface.
6. Fill the plastic syringe with water and place it against the sensor waste port (item 4,
Figure 6-2).
7. Press the syringe plunger gently to insert water into the CO-ox waste port, while
holding gauze against the end of the tubing that is connected to the
CO-ox waste exit port (item 2, Figure 6-2), to catch the obstruction and any excess
liquid.
BIOHAZARD:
practices of your workplace.
02087462 Rev. V
sample connector (item 3, Figure 6-1).
Fitting to which the reagent manifold tubing and CO-ox pump tubing are connected
(Rapidlab 1245 and 1265 systems)
Reagent manifold tubing that connects the fitting to the port on the top, left-side of the
reagent manifold (Rapidlab 1245 and 1265 systems)
Sample connector
Waste tubing connected to port on top, right-side of the reagent manifold
Two latches that secure the reagent manifold to the system
Reagent Manifold and Tubing Connections
Discard the gauze according to the established biohazard disposal
Rapidlab 1200 Operator's Guide: Troubleshooting