5. Fill the KCl reservoir with KCl fill solution:
a. Remove the reservoir cap using the hex tool and set the cap aside.
b. Partially fill the KCl reservoir.
c. Tap the front face of the sensor with your knuckle to remove any bubbles.
d. Continue filling the reservoir with KCl fill solution to the fill line.
CAUTION:
gasket and cause leaks.
6. Hand tighten the reservoir cap.
7. Clean and inspect the sensor:
a. Tap the front face of the sensor with your knuckle to remove any bubbles.
b. Wipe any excess fill solution from the exterior of the sensor using a lint-free
c. Ensure that the O-rings are in place.
8. Install the Reference sensor.
Refer to Installing the Sensor‚ page 5-31.
9. Verify sensor performance.
Refer to Verifying Sensor Performance‚ page 5-31.
02087462 Rev. V
Do not overtighten the reservoir cap. Overtightening can deform the
tissue.