Rapidlab 1200 Operator's Guide: Maintenance
b. Stand the internal electrode on its cap to drain it.
3. Fill the internal electrode compartment of the new sensor with KCl fill
solution:
a. Insert the tip of the KCl fill solution container into the internal electrode
1
2
3
Figure 5-19
b. Fill the internal electrode compartment until the KCl fill solution enters
CAUTION:
easily damaged.
4. Install the internal electrode into the internal electrode compartment and
screw the electrode into place using the hex tool.
compartment.
KCl fill solution
KCl reservoir
Internal electrode compartment
Reference Sensor Cassette and KCl Fill Solution
the KCl reservoir.
Do not touch the internal electrode wire. The wire is fragile and is
5-37
02087462 Rev. V