Rapidlab 1200 Operator's Guide: Maintenance

Flushing the Lower Connector

1. Connect the syringe tubing to the lower connector for the wash reagent.
2. Flush with a minimum of 3 mL water.
3. Flush with a minimum of 12 mL air.
4. Capture the water in an absorbent swab or equivalent.
1
2
Figure 5-27
Water collection area
Lower connector
Reagent Manifold Lower Connector
5-53