Flushing The Tubing For The Samples - Siemens Rapidlab 1200 Operator's Manual

Table of Contents

Advertisement

5-54

Flushing the Tubing for the Samples

1. Connect the syringe tubing to the connector for samples.
2. Flush the sample tubing with a minimum of 3 mL water.
3. Flush the sample tubing with a minimum of 12 mL air.
4. Collect the water in an absorbent swab or equivalent.
1
2
Figure 5-28
02087462 Rev. V
Water collection area
Sample connector
Sample Tubing
Rapidlab 1200 Operator's Guide: Maintenance