Siemens Rapidlab 1200 Operator's Manual page 429

Table of Contents

Advertisement

Rapidlab 1200 Operator's Guide: System Specifications
Table E-67
Mode
Micro-
capillary
Micro-
syringe
Table E-68
Mode
Capillary
Syringe
Micro-
capillary
Micro-
syringe
Table E-69
Mode
Capillary
Syringe
Micro-
capillary
Micro-
syringe
Table E-70
Mode
Capillary
Syringe
Micro-
capillary
Micro-
syringe
Table E-71
Mode
Capillary
Syringe
Micro-
capillary
Rapidlab 1265 Human Whole Blood Method Comparison—pCO
N
Slope
Intercept
107
1.011
-0.480
216
0.991
-0.772
Rapidlab 1265 Human Whole Blood Method Comparison—pO
N
Slope
Intercept
198
0.999
1.888
410
0.979
4.294
107
0.967
3.922
216
0.964
3.089
Rapidlab 1265 Human Whole Blood Method Comparison—Na
N
Slope
Intercept
193
0.991
1.159
387
1.003
-0.417
137
1.006
-0.603
209
1.021
-2.613
Rapidlab 1265 Human Whole Blood Method Comparison—K
N
Slope
Intercept
215
0.987
0.082
431
1.021
-0.084
143
1.007
0.005
278
1.015
-0.042
Rapidlab 1265 Human Whole Blood Method Comparison—Ca
N
Slope
Intercept
117
1.088
-0.103
259
1.108
-0.133
71
1.008
-0.016
2
RMSE
r
3.939
0.996
3.258
0.997
2
RMSE
r
Min
5.745
0.999
29.2
8.343
0.998
28.3
7.086
0.999
26.3
8.702
0.998
26.4
RMSE
r2
Min
1.506
0.994
107.6
1.528
0.995
105.4
1.532
0.995
108.1
1.521
0.995
104.2
2
RMSE
r
Min
0.183
0.997
1.35
0.179
0.997
1.02
0.182
0.997
1.37
0.169
0.997
1.06
2
RMSE
r
Min
0.092
0.994
0.52
0.090
0.995
0.49
0.050
0.998
0.52
E-69
2
Min
Max
6.0
232.5
5.9
232.3
2
Max
676.3
683.6
674.4
676.1
+
Max
180.2
182.8
189.5
187.5
+
Max
12.27
12.27
11.85
12.27
++
Max
4.59
4.86
4.35
02087462 Rev.V