Removing Clots Using The Clot-Removal Line; Inspecting The Sample Path - Siemens Rapidlab 1200 Operator's Manual

Table of Contents

Advertisement

6-24
e. Press the black spring enclosure assembly lever to reapply pressure to the sensor
10. Close the measurement module cover.
11. Close the front cover of the system.
12. Select
13. Select the Return button twice to exit Diagnostics mode.
You are prompted to calibrate the system.
14. Select
15. Replace the reagent sample port by following this procedure:
a. Select
b. Select the
16. If the reagent manifold, preheater, and reagent cartridge clot clearing procedures fail
to locate the obstruction, see Removing Clots Using the Clot-Removal Line‚
page 6-24.

Removing Clots Using the Clot-Removal Line

Use the following procedure if the clot clearing procedures in the preceding section fail to
remove the obstruction.
BIOHAZARD:
Refer to Appendix A, Protecting Yourself from Biohazards for recommended
precautions when working with biohazardous materials.
CAUTION:
without first discharging static buildup. Touch the inner surface of the
measurement module frame to discharge static buildup.

Inspecting the Sample Path

1. Remove the reagent cartridge and the wash cartridge:
a. Select
b. Remove both the cartridges.
c. Leave the doors open.
02087462 Rev. V
assembly.
.
Stop
to calibrate the system.
Yes
Status > Maintenance > Replace Sample Port
.
Continue
Follow the instructions in the video to replace the sample port.
Wear personal protective equipment. Use universal precautions.
Do not remove or return the sensors to the measurement module
Status > Diagnostics > Cartridges > Eject Both Cartridges
Rapidlab 1200 Operator's Guide: Troubleshooting
.
.