Siemens Rapidlab 1200 Operator's Manual page 381

Table of Contents

Advertisement

Rapidlab 1200 Operator's Guide: System Specifications
The following tables summarize the results of the Rapidlab 1245 system human
whole blood recovery precision testing.
Table E-18
Level Mode
1
2
3
4
5
6
1
2
3
4
5
6
1
2
3
4
5
6
1
2
3
4
5
6
Rapidlab 1245 Recovery and Precision Testing—pH
n
WRSD
Capillary
21
0.004
Capillary
24
0.003
Capillary
21
0.004
Capillary
24
0.003
Capillary
21
0.002
Capillary
24
0.002
Syringe
48
0.006
Syringe
48
0.005
Syringe
48
0.004
Syringe
48
0.003
Syringe
48
0.002
Syringe
48
0.002
Micro-
24
0.011
capillary
Micro-
24
0.011
capillary
Micro-
24
0.006
capillary
Micro-
24
0.009
capillary
Micro-
24
0.008
capillary
Micro-
24
0.021
capillary
Micro-
24
0.007
syringe
Micro-
24
0.004
syringe
Micro-
24
0.003
syringe
Micro-
24
0.003
syringe
Micro-
24
0.004
syringe
Micro-
24
0.003
syringe
*
Observed
Expected
7.032
7.032
7.205
7.202
7.347
7.354
7.409
7.410
7.513
7.534
7.664
7.630
7.037
7.032
7.213
7.202
7.347
7.354
7.422
7.410
7.546
7.534
7.626
7.630
7.046
7.032
7.197
7.202
7.366
7.354
7.413
7.410
7.540
7.534
7.624
7.630
7.033
7.032
7.193
7.202
7.355
7.354
7.413
7.410
7.548
7.534
7.619
7.630
E-21
Recovery
WRCV
100.00
0.06
100.04
0.04
99.90
0.06
99.99
0.04
99.72
0.03
100.45
0.02
100.07
0.09
100.15
0.07
99.92
0.05
100.16
0.03
100.15
0.03
99.95
0.03
100.20
0.16
99.92
0.15
100.17
0.08
100.03
0.13
100.08
0.10
99.92
0.27
100.01
0.11
99.86
0.06
100.01
0.05
100.03
0.04
100.18
0.05
99.85
0.05