Siemens Rapidlab 1200 Operator's Manual page 421

Table of Contents

Advertisement

Rapidlab 1200 Operator's Guide: System Specifications
Table E-59
Level Mode
5
6
1
2
3
4
5
6
1
2
3
4
5
6
*
WRSD = within-run standard deviation
Table E-60
Level Mode
1
2
3
4
5
6
1
2
Rapidlab 1265 Recovery and Precision Testing—Cl
N
WRSD
Syringe
42
0.38
Syringe
45
0.98
Micro-
12
0.68
capillary
Micro-
12
0.43
capillary
Micro-
12
0.43
capillary
Micro-
12
0.53
capillary
Micro-
11
0.31
capillary
Micro-
12
1.22
capillary
Micro-
24
0.53
syringe
Micro-
24
0.31
syringe
Micro-
24
0.37
syringe
Micro-
24
0.78
syringe
Micro-
24
0.65
syringe
Micro-
24
1.63
syringe
Rapidlab 1265 Recovery and Precision Testing—Ca
N
WRSD
Capillary
18
0.009
Capillary
21
0.015
Capillary
18
0.004
Capillary
21
0.023
Capillary
18
0.026
Capillary
21
0.039
Syringe
40
0.011
Syringe
45
0.009
*
Observed
Expected
110
111
146
146
80
80
98
98
98
100
104
105
116
117
141
141
81
81
95
96
98
99
106
106
117
116
145
144
*
Observed
Expected
0.57
0.58
1.26
1.26
1.34
1.36
1.90
1.93
2.29
2.36
4.08
4.09
0.55
0.58
1.25
1.26
E-61
-
Recovery
WRCV
99.7
0.3
100.6
0.7
100.1
0.9
100.1
0.4
98.8
0.4
98.8
0.5
99.5
0.3
99.3
0.9
100.1
0.7
99.5
0.3
98.8
0.4
99.9
0.7
101.0
0.6
101.1
1.1
++
Recovery
WRCV
97.0
1.6
100.2
1.2
98.4
0.3
98.7
1.2
97.1
1.1
99.6
1.0
94.6
2.0
99.3
0.7
02087462 Rev.V

Advertisement

Table of Contents
loading

Table of Contents