Rapidlab 1200 Operator's Guide: System Specifications
The following tables summarize the results of the Rapidlab 1240 system recovery
and precision testing:
Table E-11
Level Mode
1
2
3
4
5
6
1
2
3
4
5
6
1
2
3
4
5
6
1
2
3
4
5
6
Rapidlab 1240 Recovery and Precision Testing—pH
n
WRSD
Capillary
24
0.004
Capillary
24
0.002
Capillary
20
0.006
Capillary
24
0.004
Capillary
24
0.002
Capillary
24
0.004
Syringe
48
0.004
Syringe
48
0.005
Syringe
47
0.003
Syringe
48
0.003
Syringe
48
0.004
Syringe
47
0.003
Micro-
24
0.01
capillary
Micro-
24
0.013
capillary
Micro-
24
0.011
capillary
Micro-
24
0.012
capillary
Micro-
24
0.007
capillary
Micro-
24
0.006
capillary
Micro-
24
0.004
syringe
Micro-
24
0.002
syringe
Micro-
24
0.006
syringe
Micro-
24
0.003
syringe
Micro-
24
0.005
syringe
Micro-
24
0.003
syringe
*
Observed
Expected
7.042
7.032
7.204
7.202
7.349
7.354
7.409
7.41
7.516
7.534
7.662
7.63
7.034
7.032
7.209
7.202
7.353
7.354
7.418
7.41
7.545
7.534
7.621
7.63
7.03
7.032
7.204
7.202
7.359
7.354
7.403
7.41
7.538
7.534
7.629
7.63
7.031
7.032
7.197
7.202
7.356
7.354
7.418
7.41
7.556
7.534
7.628
7.63
E-13
Recovery
WRCV
100.15
0.05
100.02
0.03
99.94
0.09
99.99
0.05
99.76
0.03
100.42
0.05
100.04
0.06
100.09
0.07
99.99
0.05
100.11
0.04
100.15
0.05
99.88
0.04
99.97
0.14
100.02
0.18
100.07
0.15
99.91
0.16
100.05
0.09
99.98
0.08
99.98
0.05
99.92
0.03
100.03
0.08
100.1
0.05
100.28
0.06
99.97
0.04
02087462 Rev.V