Using The Recall-Levey-Jennings Screens; Viewing The Recall-Levey-Jennings Graph Screen - Siemens Rapidlab 1200 Operator's Manual

Table of Contents

Advertisement

Rapidlab 1200 Operator's Guide: Quality Control
To print the Recall–QC Statistics data, select the
contains the following additional information:
parameter labels and statistics for all parameters
measured and target values for each parameter
the difference between the measured and target values, in percentage
if a parameter exceeds the RiliBÄK range, text indicates the parameter has
failed
Using the Recall–Levey-Jennings Screens
The Recall–Levey-Jennings screens allow you to select a particular lot of QC
material and a parameter and then to view a month's worth of data as a
Levey-Jennings graph.
NOTE:
Levey-Jennings graph.
1. At the user interface, select
2. At the Recall screen, select QC, and then select
3. At the Recall–Levey-Jennings Graph screen, select the lot number for the QC
material.
Use the up and down scroll arrows to navigate the list. The list is in
alphanumeric order, with the AutomaticQC lots listed first as AQC1, AQC2,
and so on.
4. Select a month.
The current month is uppermost of the three buttons. The 2 previous months
are next in order.
5. Select a parameter.
6. Select
This displays the Recall Levey-Jennings Graph screen.
Viewing the Recall–Levey-Jennings Graph Screen
The Levey-Jennings graph displays a visual representation of the performance of
the selected parameter relative to QC measurements for up to a 31-day period of a
single month.
Select the
Levey-Jennings graph to an external printer.
If you want more than one month of data, select and print one month and then
select and print another month.
Under the banner, the Recall–Levey-Jennings Graph screen displays the lot
information for the currently displayed lot of QC material:
parameter label for the data displayed in the graph
You must have defined range values for the system to generate a
.
Continue
button to print the Levey-Jennings graph. You can only print the
Print
Print
.
Recall
Levey-Jennings Graph
button. The printed report
.