Siemens Rapidlab 1200 Operator's Manual page 400

Table of Contents

Advertisement

E-40
Table E-38
Level Mode
1
2
3
4
5
6
1
2
3
4
5
6
1
2
3
4
5
6
1
2
3
4
5
6
*
WRSD = within-run standard deviation
02087462 Rev.V
Rapidlab 1260 Recovery and Precision Testing—K
N
WRSD
Capillary
21
0.096
Capillary
24
0.046
Capillary
24
0.034
Capillary
23
0.107
Capillary
21
0.084
Capillary
23
0.163
Syringe
45
0.078
Syringe
45
0.024
Syringe
45
0.026
Syringe
45
0.052
Syringe
45
0.116
Syringe
45
0.118
Micro-
12
0.078
capillary
Micro-
13
0.027
capillary
Micro-
12
0.024
capillary
Micro-
12
0.042
capillary
Micro-
11
0.073
capillary
Micro-
12
0.190
capillary
Micro-
22
0.088
syringe
Micro-
24
0.031
syringe
Micro-
24
0.019
syringe
Micro-
24
0.044
syringe
Micro-
20
0.159
syringe
Micro-
18
0.269
syringe
Rapidlab 1200 Operator's Guide: System Specifications
*
Observed
Expected
1.21
1.18
2.56
2.63
4.19
4.12
6.05
5.74
7.88
7.99
11.55
11.31
1.19
1.18
2.64
2.63
4.12
4.12
5.70
5.74
8.18
7.99
11.69
11.31
1.45
1.15
2.84
2.82
3.72
3.77
5.57
5.71
7.58
8.05
11.86
11.97
1.29
1.15
2.89
2.82
3.73
3.77
5.61
5.71
7.97
8.05
11.90
11.97
+
Recovery
WRCV
102.2
8.0
97.3
1.8
101.7
0.8
105.2
1.8
98.6
1.1
102.1
1.4
100.5
6.5
100.1
0.9
100.1
0.6
99.2
0.9
102.3
1.4
103.4
1.0
126.1
5.3
100.8
1.0
98.4
0.7
97.6
0.7
94.1
1.0
99.1
1.6
112.5
6.8
102.6
1.1
98.7
0.5
98.2
0.8
99.0
2.0
99.4
2.3

Advertisement