02087462 Rev.V
Rapidlab 1265 Recovery and Precision Testing—K
N
WRSD
Micro-
11
0.063
capillary
Micro-
12
0.035
capillary
Micro-
12
0.066
capillary
Micro-
12
0.034
capillary
Micro-
12
0.085
capillary
Micro-
11
0.242
capillary
Micro-
25
0.091
syringe
Micro-
24
0.034
syringe
Micro-
24
0.016
syringe
Micro-
24
0.058
syringe
Micro-
24
0.134
syringe
Micro-
24
0.280
syringe
Rapidlab 1265 Recovery and Precision Testing—Cl
N
WRSD
Capillary
21
0.38
Capillary
21
0.38
Capillary
21
0.38
Capillary
21
0.38
Capillary
21
0.38
Capillary
20
0.38
Syringe
45
0.38
Syringe
45
0.38
Syringe
45
0.42
Syringe
43
0.69
Rapidlab 1200 Operator's Guide: System Specifications
*
Observed
Expected
1.27
1.15
2.77
2.82
3.85
3.77
5.63
5.71
7.72
8.05
12.28
11.97
1.15
1.15
2.78
2.82
3.77
3.77
5.71
5.71
8.22
8.05
12.38
11.97
*
Observed
Expected
72
73
90
91
97
98
105
106
111
111
148
148
70
71
90
91
99
99
106
105
+
Recovery
WRCV
110.4
4.9
98.0
1.3
101.9
1.7
98.6
0.6
95.9
1.1
102.6
2.0
99.6
7.9
98.7
1.2
99.8
0.4
100.1
1.0
102.1
1.6
103.4
2.3
-
Recovery
WRCV
99.2
0.5
99.6
0.5
99.3
0.3
98.5
0.6
99.8
0.4
100.0
1.4
98.0
1.4
99.7
0.5
100.2
0.4
100.3
0.7