Understanding System Behavior With Custom Panels - Siemens Rapidlab 1200 Operator's Manual

Table of Contents

Advertisement

Rapidlab 1200 Operator's Guide: Operating the System

Understanding System Behavior with Custom Panels

When you select a custom panel, the system responds as follows:
The system disables any functions that you cannot use with custom panels.
If you select a custom panel that contains the tHb parameter and then select
Microsample, the tHb parameter is removed from the custom panel while
performing that analysis.
If you select a custom panel that does not contain the pO
parameters and then select Microsample, the system does not perform the gas
measurement phase of the microsample measurement.
The Analysis screen does not display custom panels under some conditions:
you did not define custom panels in Setup
if you select a sample type other than a patient sample type, such as QC,
AQC, or High G/L
if you select pH only, pH Glu Lac only, or tHb only modes
2-21
and pCO
2
2
02087462 Rev. V

Advertisement

Table of Contents
loading

Table of Contents