Appendix E: System Specifications; Viewing Parameter Measurements - Siemens Rapidlab 1200 Operator's Manual

Table of Contents

Advertisement

Appendix E: System Specifications

The following table lists the specifications for the Rapidlab 1200 systems.
Table E-1
Property
Ambient operating temperature
Ambient temperature change
Ambient operating relative humidity
Ambient shipping temperature
Ambient storage temperature
Power rating
Voltage requirements
Ambient operating barometric
pressure
System dimensions, base model
*
Height with screen fully upright

Viewing Parameter Measurements

The following table lists the units, reporting ranges, and display resolutions for the
pH and blood gas parameters measured by the Rapidlab 1200 system.
Table E-2
Parameter
pH
+
H
pCO
2
pO
2
pAtm
Rapidlab 1200 System Specifications
pH and Blood Gas Parameters
Units
pH
nmol/L
mmHg
kPa
mmHg
kPa
mmHg
kPa
Specification
15–32°C
less than 4°C per hour
5–90%, non-condensing
-25–45°C
4–45°C
150VA
100V/240V (85V to 264V autoranging)
50/60Hz
523–800 mmHg (69.7–106.7 kPa)
height 61.0 cm (24 inches)
width 53.5 cm (21 inches)
depth 58.5 cm (23 inches)
weight 29.5 kg (65 lbs)
Reporting Range
6.000–8.000
10.0–1000.0
5.0–250.0
0.67–33.33
0–800
0.0–106.67
523–800
69.7–106.6
*
Resolution
0.001
0.1
0.1
0.01
0.1
0.01
1
0.1
02087462 Rev.V

Advertisement

Table of Contents
loading