Troubleshooting Failed Calibrations

Use this section to help determine causes and corrective actions for failed calibrations.
Error
The parameter
has a single
line through it.
The parameter
has 2 crossed
lines through
it.

Troubleshooting Checklist for Failed Calibration

Perform the solutions in this checklist to help clear problems with failed calibrations:
If D codes also exist for reagent problems such as D23, correct these problems first.
Deproteinize the sample path if not done within the last week.
If the sensor was newly installed or if maintenance was recently performed, allow
sufficient time for the sensor to equilibrate.
Perform additional calibrations.
For pH, Na
check the solution level in the reference sensor and look for salt in the vent hole,
bubbles in the sensor, or leaks.
02087462 Rev. V
Rapidlab 1200 Operator's Guide: Troubleshooting
Probable
Cause
Corrective Action
The parameter
For recommendations to resolve failed calibrations,
failed
refer to Troubleshooting Checklist for Failed
calibration.
Calibration‚ page 6-4.
View the events log to determine the event causing
the parameter to fail and follow the appropriate
corrective action for that event.
The parameter
Perform a 2-point calibration to make the
was turned off
parameter available again.
in Setup and
then turned on
again
The parameter
View the events log to determine the event causing
failed
the parameter to fail and follow the appropriate
calibration and
corrective action for that event.
is unlikely to
become
available with
further
calibrations.
+
+
++
, K
, Ca
, and Cl¯ sensors, if more than 2 sensors have excessive drift,