Siemens Rapidlab 1200 Operator's Manual page 425

Table of Contents

Advertisement

Rapidlab 1200 Operator's Guide: System Specifications
Table E-62
Level Mode
1
2
3
4
5
1
2
3
4
5
1
2
3
4
5
*
WRSD = within-run standard deviation
Rapidlab 1265 Recovery and Precision Testing—Lactate
N
WRSD
Micro-
24
0.06
syringe
Micro-
24
0.06
syringe
Micro-
24
0.14
syringe
Micro-
24
0.23
syringe
Micro-
24
0.23
syringe
pH, Glu,
21
0.05
Lac
Capillary
pH, Glu,
21
0.07
Lac
Capillary
pH, Glu,
15
0.06
Lac
Capillary
pH, Glu,
21
0.13
Lac
Capillary
pH, Glu,
15
0.25
Lac
Capillary
pH, Glu,
41
0.03
Lac
Syringe
pH, Glu,
45
0.08
Lac
Syringe
pH, Glu,
45
0.11
Lac
Syringe
pH, Glu,
42
0.33
Lac
Syringe
pH, Glu,
45
0.34
Lac
Syringe
*
Observed
Expected
1.2
1.2
3.3
3.6
5.9
6.4
9.7
10.8
11.9
13.9
0.9
0.9
2.8
2.8
4.9
5.3
10.5
12.1
14.2
17.4
0.9
0.8
2.5
2.7
5.5
6.0
10.8
12.5
14.8
17.9
E-65
Recovery
WRCV
98.6
5.1
90.6
2.0
93.0
2.3
89.4
2.3
85.9
2.0
109.7
5.1
98.5
2.6
92.9
1.3
86.4
1.2
81.5
1.7
106.5
3.1
93.5
3.2
91.4
1.9
86.9
3.1
82.8
2.3
02087462 Rev.V