D6 Excessive Noise: Ph Na K Ca Cl; D6 Excessive Noise: Glu Lac - Siemens Rapidlab 1200 Operator's Manual

Table of Contents

Advertisement

Rapidlab 1200 Operator's Guide: Troubleshooting
D6 Excessive Noise: pH Na
Problem
The sensor signal
has excess noise.
The patient sample
result for the
affected parameter
is reported as
question result
(-----?).

D6 Excessive Noise: Glu Lac

Problem
The sensor signal
has excess noise.
The patient sample
result for the
affected parameter
is question result
(-----?).
+
+
K
Ca
Corrective Action
Correct any reagent problems such as D23 codes before you
attempt to correct this D code.
Ensure that the measurement module door was not opened
during analysis.
If door was opened, repeat the sample analysis.
Perform a successful calibration (1-point or 2-point).
Refer to Performing Manual Calibrations‚ page 3-4.
If a D6 occurs during sample analysis, repeat the sample.
Deproteinize the sample path and condition the sensors.
Refer to Deproteinizing the Sample Path‚ page 5-4.
Check the solution level in the affected measurement sensor
and look for bubbles.
+
++
The K
, Ca
, and Cl¯ sensors should be nearly full. The pH
+
and Na
sensors should be full. If the solution level is low,
refill the sensor.
If a D6 occurs for more than one sensor, check the solution
level in the reference sensor and look for salt in the vent hole
and bubbles in the sensor.
Clean and refill the reference sensor as needed.
Check the sensor contacts and measurement block for salt
buildup.
Check behind the sensors for fluid leaks.
If salt or leaks are present, clean the sensor with water, dry,
reinstall, and perform a 2-point calibration.
Refer to Performing Manual Calibrations‚ page 3-4.
If the problem persists, replace the sensor.
If the problem persists, call for technical assistance.
Corrective Action
Correct any reagent problems such as D23 codes before you
attempt to correct this D code.
Perform a successful 1-point or 2-point calibration.
Refer to Performing Manual Calibrations‚ page 3-4.
If a D6 occurs during sample analysis, repeat the sample.
If the problem persists, replace the sensor.
If the problem persists, call for technical assistance.
++
Cl¯
02087462 Rev. V
6-57