D37 Codes

D37 Cartridge Eject Error: 1

Problem
The eject
mechanism fails to
move to the correct
position.

D37 Cartridge Eject Error: 2, 3

Problem
The system fails to
open the cartridge
door during
cartridge
replacement.
2 reagent cartridge
3 wash cartridge
D38–D60 Codes

D38 Temp Error: 1-12

Problem
A component in the
temperature control
system failed or the
ambient
temperature is
beyond the
acceptable
operating
conditions.
02087462 Rev. V
Rapidlab 1200 Operator's Guide: Troubleshooting
Corrective Action
1.
Select
Diagnostics > Cartridge > Eject W Cartridge
.
Eject R Cartridge
2.
Select
Status > Shutdown
3.
Wait 10 seconds, and then turn the system on.
4.
If the problem persists, call for technical assistance.
Corrective Action
1.
Ensure that nothing is blocking the door.
2.
Select
Diagnostics > Cartridge > Eject W Cartridge
.
Eject R Cartridge
3.
If the problem persists, call for technical assistance.
Corrective Action
Check the Gnd sensor and ensure that it is aligned with the
contacts.
If the problem persists, continue with the following:
The system attempts to re-enable the temperature control every
15 minutes and re-evaluates the error condition a maximum of 3
times. This temperature control process can take up to 3.5 hours.
To determine the temperatures in the instrument, select
Diagnostics > Temperature/pAtm
Ensure the ambient temperature is within 15
Ensure that the front cover is closed.
Wait while the system attempts to correct the problem.
If the problem persists, select
Wait 10 seconds, and then turn the system on.
If the problem persists, call for technical assistance.
.
.
°
to 32°C.
.
Status > Shutdown