Entering Customized QC Range Limits

1. At the Setup screen, select
2. If performing AQC, select
If performing RQC, select
3. Select the QC range to view.
The screen for the QC range you selected displays.
4. Select a parameter.
5. Enter a low limit, a high limit, or both, in the Low and High boxes, for the
selected parameter.
If in RiliBÄK mode, use the following equations to determine low and high
values:
Range Offset from Manufacturing Target =
Manufacturing Target + (Manufacturing Target * Desired Deviation from
Target)
QC Low Value =
(Manufacturing Target – Range Offset from Manufacturing Target)
QC High Value =
(Manufacturing Target + Range Offset from Manufacturing Target)
6. Repeat steps 4-5 as needed, for additional parameters.
7. Select

Setting Up Required QC

Use the Required QC screen to set up a required QC sample schedule. Use the
barcode scanner and RapidQC controls to perform Required QC analysis.
NOTE:
schedule. Refer to Setting up Required QC Ranges‚ page 8-4. You cannot analyze
the scheduled controls until you enter the lot information.
1. At the Setup screen, select
The screen displays the Required QC Schedule with the current schedule for
Required QC analysis, if defined.
NOTE:
that were previously scheduled.
2. Select
The system displays the Intervals screen.
3. Select the number of hours for the interval of the Required QC analysis.
For example, select 8 to have an 8-hour window for performing scheduled
controls.
.
Continue
Ensure that you enter information for each lot of the controls you
When you change intervals, the system clears any times and levels
.
Intervals
.
QC
AutomaticQC Ranges
Required QC Ranges
QC > Required QC Schedule
.
.
.