Rapidlab 1200 Operator's Guide: System Configuration
6. At the Required QC Ranges screen, select

Editing Required QC Ranges

You can change the range values of each parameter in the selected level. You can
also convert a trial lot to an active lot.
1. At the Setup screen, select
2. Select an available level button to view or edit the ranges for that level:
Selecting the level button displays the Range Entry screen displaying the
ranges defined for each parameter for that level.
3. Select the parameter you want to edit and then select
4. To clear the field, select
5. Use the numeric buttons to enter new range values.
6. Repeat steps 3 through 5 for each parameter you need to edit.
7. If the parameter is a trial lot and it should be an active lot, select
The Set Active button is only displayed when the lot is a trial lot.
8. Use the print button to print the values selected for this level.
9. Select

Setting up AutomaticQC Schedule

Use this screen to select the days and times for AutomaticQC analysis and select
the controls to analyze at the scheduled times.
1. At the Setup screen, select
If a schedule is already defined, the AutomaticQC Schedule screen displays
the currently defined schedule.
2. Select each button for the level you want to schedule.
a. Use the arrow keys to navigate to a place in the grid indicating the day of
b. Select one or more buttons for the level or levels to schedule.
3. Turn AutomaticQC analysis on:
a. Select
b. Select
c. Select
Clear
.
Save
the week and time period for the level or levels.
A check mark on the button means that you scheduled that level. Select
the button again to remove the check mark.
QC > QC Options > AutomaticQC
if you want the system to repeat QC analysis when a
Repeat
parameter fails the first analysis.
.
Save
to return to the Setup menu.
Save
QC > Required QC Ranges
.
QC > AutomaticQC Schedule
.
8-5
.
or
.
Low
High
.
Set Active
.