Rapidlab 1200 Operator's Guide: Troubleshooting
b. If you detect an obstruction, fill a syringe with water, place the syringe
1
2
Figure 6-11
CAUTION:
ground/temperature, pH, and reference sensors.
c. For the sample ground/temperature sensor, pH sensor, and reference
1
2
Figure 6-12
against the sample path and inject water through the sample path.
Sensor
Syringe
Injecting Water into the Gas Sensor Sample Path
Use the clot-removal line carefully to avoid damage to the sample
sensor, push the clot-removal line through the sample path to remove
obstructions.
Sensor cartridge
Clot-removal line
Clot-Removal Line in Sensor Sample Path
6-27
02087462 Rev. V