Final procedures

Service PC and measuring instruments

Saving the configuration file

NOTE
Main menu:
1. Select <Configuration> <Save config file>.
2. Press <F2> to save data.

Reading the exposure counter

Main menu:
1. Select <Service> <Show exposure counter>.
2. Enter the current counter reading in the certificate.

Deleting the error memory

Main menu:
1. Select <Service> <Delete error buffer>.
2. Delete the error memory with <Y>.

Saving the programmed values

Main menu:
1. Select <Backup> <Copy installation area to floppy> <All>.
2. Leave the program with <F10> when the backup is ready.
NOTE
Siemens AG
Medical Solutions
Make sure that the backup floppy is inserted.
Keep the backup floppy with other documentation for the
MAMMOMAT in question.