Siemens Rapidlab 1200 Operator's Manual page 418

Table of Contents

Advertisement

E-58
Table E-56
Level Mode
1
2
3
4
5
6
7
8
9
*
WRSD = within-run standard deviation
Table E-57
Level Mode
1
2
3
4
5
6
1
2
3
4
5
6
1
2
02087462 Rev.V
Rapidlab 1265 Recovery and Precision Testing—pO
N
WRSD
Micro-
24
0.41
syringe
Micro-
24
0.22
syringe
Micro-
24
0.97
syringe
Micro-
24
1.53
syringe
Micro-
24
1.17
syringe
Micro-
24
2.52
syringe
Micro-
24
2.74
syringe
Micro-
24
4.31
syringe
Micro-
24
8.16
syringe
Rapidlab 1265 Recovery and Precision Testing—Na
N
WRSD
Capillary
21
0.18
Capillary
21
0.19
Capillary
21
0.42
Capillary
21
0.84
Capillary
21
0.25
Capillary
21
1.06
Syringe
45
0.41
Syringe
45
0.40
Syringe
45
0.36
Syringe
43
0.33
Syringe
42
0.25
Syringe
45
0.65
Micro-
12
0.56
capillary
Micro-
12
0.27
capillary
Rapidlab 1200 Operator's Guide: System Specifications
*
Observed
Expected
28.0
28.6
48.3
49.5
84.1
85.6
105.5
106.1
149.9
149.7
200.5
199.9
263.0
249.6
376.9
378.4
640.8
643.4
*
Observed
Expected
113.6
111.2
129.6
129.4
134.0
134.4
142.1
141.5
147.4
148.6
175.3
179.8
113.5
111.2
129.0
129.4
133.6
134.4
140.1
141.5
148.5
148.6
173.7
179.8
110.0
109.9
129.9
130.1
2
Recovery
WRCV
98.0
1.5
97.5
0.5
98.3
1.1
99.5
1.5
100.1
0.8
100.3
1.3
105.4
1.0
99.6
1.1
99.6
1.3
+
Recovery
WRCV
102.1
0.2
100.1
0.1
99.7
0.3
100.4
0.6
99.2
0.2
97.5
0.6
102.1
0.4
99.6
0.3
99.4
0.3
99.0
0.2
99.9
0.2
96.6
0.4
100.1
0.5
99.9
0.2

Advertisement