Rapidlab 1200 Operator's Guide: Maintenance
Maintaining the Sensors
Performing regular maintenance on the sensors is important to the life of the
sensors. This section contains the following information:
•
Replacing the sensors
•
Performing reference sensor maintenance
•
Refilling the sensors
Replacing the Sensors
You need the following materials for this task:
•
Appropriate sensor
•
Appropriate sensor fill solution
•
Reference sensor replacement kit
•
Hex tool
•
Lint-free tissue
BIOHAZARD:
Refer to Appendix A, Protecting Yourself from Biohazards for recommended
precautions when working with biohazardous materials.
CAUTION:
without first discharging static buildup. Touch the inner surface of the
measurement module frame to discharge static buildup.
Preparing the Sensors
•
Equilibrate new Glucose and Lactate biosensors at room temperature
(18° to 25°C) for at least 1 hour before installing.
•
Keep the biosensors in their foil packages while they are equilibrating.
Wear personal protective equipment. Use universal precautions.
Do not remove or return the sensors to the measurement module