Rapidlab 1200 Operator's Guide: Maintenance

Maintaining the Sensors

Performing regular maintenance on the sensors is important to the life of the
sensors. This section contains the following information:

Replacing the sensors

Performing reference sensor maintenance
Refilling the sensors
Replacing the Sensors
You need the following materials for this task:
Appropriate sensor
Appropriate sensor fill solution
Reference sensor replacement kit
Hex tool
Lint-free tissue
BIOHAZARD:
Refer to Appendix A, Protecting Yourself from Biohazards for recommended
precautions when working with biohazardous materials.
CAUTION:
without first discharging static buildup. Touch the inner surface of the
measurement module frame to discharge static buildup.

Preparing the Sensors

Equilibrate new Glucose and Lactate biosensors at room temperature
(18° to 25°C) for at least 1 hour before installing.
Keep the biosensors in their foil packages while they are equilibrating.
Wear personal protective equipment. Use universal precautions.
Do not remove or return the sensors to the measurement module
5-29
02087462 Rev. V