6-2
Error
02087462 Rev. V
Rapidlab 1200 Operator's Guide: Troubleshooting
Probable
Cause
Corrective Action
Parameter
Perform the following steps until the system is
failed
ready for use:
AutomaticQC
•
Perform up to two 2-point calibrations.
analysis
•
If the parameter fails the 2-point calibrations,
because the
view the events log for D codes and messages.
sensor or the
•
Replace the sensor if necessary.
cartridge is not
•
If the parameter passed the 2-point
working
calibrations, perform AutomaticQC analysis
correctly.
for the failed level of control.
•
If the parameter fails AutomaticQC analysis,
analyze a QC sample for the failed level using
QC material from an ampule.
•
If the parameter fails the ampule QC analysis,
review the events log for D codes and
messages.
•
If the parameter passes the ampule QC
analysis, replace the AutomaticQC cartridge.
•
Verify that the target ranges entered for the
The target
control are correct.
ranges for one
or more
If you specify target ranges that are narrower
than those provided with the control,
parameters are
parameters may fail QC analysis.
too narrow.
•
Analyze QC for the failed level.
Refer to Performing Unscheduled QC Sample
Analysis‚ page 4-4.
•
If the results are within range, restore the
parameter.
Refer to Restoring Parameters‚ page 4-12.
•
For AutomaticQC samples, analyze the failed
level.
Refer to Performing Unscheduled QC Sample
Analysis‚ page 4-4.
If the results are within range, the system
restores the parameter.
•
If analysis fails again, review the events log for
D codes and messages.