Rapidlab 1200 Operator's Guide: System Specifications
Table E-13
Level Mode
5
6
7
8
9
1
2
3
4
5
6
7
8
9
*
WRSD = within-run standard deviation
Method Comparison with Human Whole Blood Samples
For testing syringe and capillary modes, blood was collected in heparinized
vacuum tubes. It was tonometered at 37°C and/or adjusted chemically. Several
levels of tHb were prepared by separating the red cells from the plasma and
recombining appropriately. Specimens were assayed on Rapidlab 1240 and
Rapidlab 865 systems.
Rapidlab 1240 Recovery and Precision Testing—pO
n
WRSD
Micro-
12
0.66
capillary
Micro-
12
1.72
capillary
Micro-
12
2.09
capillary
Micro-
12
2.71
capillary
Micro-
12
6.27
capillary
Micro-
24
0.17
syringe
Micro-
24
0.33
syringe
Micro-
24
0.48
syringe
Micro-
24
0.85
syringe
Micro-
24
5.95
syringe
Micro-
24
3.7
syringe
Micro-
24
3.47
syringe
Micro-
24
6.07
syringe
Micro-
24
7.59
syringe
*
Observed
Expected
159.9
149.8
199.9
199.5
265
249.7
371
377.6
632.5
649.3
30.4
28.3
54
50.1
90.2
85.9
104.6
107.4
155.4
150.3
201.9
198.2
261.6
250.5
374.5
375.1
642.1
651.3
E-17
2
Recovery
WRCV
106.7
0.4
100.2
0.9
106.1
0.8
98.2
0.7
97.4
1
107.2
0.6
107.8
0.6
105
0.5
97.4
0.8
103.4
3.8
101.9
1.8
104.4
1.3
99.8
1.6
98.6
1.2
02087462 Rev.V