Siemens Rapidlab 1200 Operator's Manual page 238

Table of Contents

Advertisement

6-32
b. Rotate the gray wheel towards you, pull the tubing through the fluid detector, and
c. Push the new tubing through the fluid detector.
d. Connect the tubing to the steel measurement module connector and to the sample
3. Remove obstructions from the CO-ox waste tubing:
a. Disconnect the tubing from the steel fluid connector and the connector in the
1
Figure 6-16
b. Push a clot-removal line through the tubing, opposite the direction of sample flow.
c. Reconnect the CO-ox waste tubing.
4. Remove obstructions from the waste path in the reagent manifold:
a. Disconnect the CO-ox pump tubing from the reagent manifold tubing at the
02087462 Rev. V
remove the tubing from the steel measurement module connector.
chamber connector.
Use a clamp or tweezers to help you attach the tubing.
pump tubing.
Steel fluid connector
CO-ox Sample Path
fitting.
Rapidlab 1200 Operator's Guide: Troubleshooting