System Operation
System Settings
Start
12
/ 32
Object-dependent system settings
The agreement of the actuation direction of the operating element with the
movement of the X-ray system (stand with tube assembly and primary collimator,
image receptor unit with image intensifier and spotfilm device) depends upon the
selected object setting.
❏ Left LED lights up: "Object setting oriented to the monitor" is selected
❏ Right LED lights up: "Object setting oriented to the system" is selected
Only possible from the system remote control console for the following move-
ments:
– Moving the X-ray system longitudinally
– Moving the tabletop longitudinally/transversely
Start positions
Automatically determined system settings are performed with the Start button
depending upon the selected operating mode (tomography, Periscanning,
Peristepping, scanning technique).
LED flashes: the approved system settings are not yet concluded.
LED does not light up: the approved system settings are concluded.
General notes on error messages
The system is constantly monitored internally during operation.
When a fault is detected, the system is blocked and the fault is displayed as error
in the message line of the integrated generator control console and stored.
Cancelling blocking:
◆ Briefly press the exposure release button on the generator ON/OFF console
up to the 1st pressure point and let go again. Don't press more!
– The fault message is cancelled as a rule
Further operation of the system may be only under increased attention of the
user.
However, if an error message is still displayed, there is an error.
◆ Call the Siemens Uptime Service
AXD3-340.620.13.01.02
Operator Manual