Reference Methods

The following reference or comparative methods were used to generate data for the
Rapidlab 1200 systems.
Analyte
pH
pCO
2
+
Na
and K
-
Cl
++
Ca
Glucose
Lactate
tHb
FO
Hb
2
FHHb
FCOHb
FMetHb
02087462 Rev.V
Reference Method
Internal ISE method used.
and pO
Tonometered whole blood. Refer to CLSI document C46-A.
2
used are traceable to NIST Certified Reference Material SRM 1701.
Internal ISE method used.
+
Coulometric reference method.
Internal ISE method used.
The hexokinase/glucose-6-phosphate dehydrogenase method
described in CLSI document RS1-A.
Internal method using lactate oxidase assay.
Cyanmethemoglobin Reference Method per Clinical and Laboratory
Standards Institute (CLSI), approved reference procedure
spectrophotometer.
Tonometry, where whole blood samples are tonometered with 95%
O
, and 5% CO
2
Tonometry, where whole blood samples are tonometered with 95%
N
, and 5% CO
2
Reduced gas Chromatography.
Modified Evelyn-Malloy
Rapidlab 1200 Operator's Guide: System Specifications
4
.
2
.
2
6
7
method utilizing a spectrophotometer.
2
Gases
5
using a