Full 2-Port Calibration Using The 2-Port Ecal Module - Agilent Technologies ENA Series User Manual

Rf network analyzer
Hide thumbs Also See for ENA Series:
Table of Contents

Advertisement

Step 6. Perform a 1-port calibration.
Softkey
Port 1
Port 2
Port 3
Port 4
*1.Only with Options 313, 314, 413, and 414.
*2.Only with Options 413 and 414.

Full 2-Port Calibration Using the 2-Port ECal Module

Follow the procedure below to perform a full 2-port calibration using the 2-port ECal
module.
Step 1. Connect the USB port on the ECal module with the USB port on the E5070A/E5071A
using a USB cable. This connection may be done with the E5070A/E5071A power on.
Step 2. Connect port A and port B on the ECal module to the test ports that need to be calibrated.
Figure 4-19
Connecting the ECal Module (Full 2-Port Calibration)
Step 3. Press
Step 4. Press
Step 5. To enable isolation calibration, press
Step 6. Press
this key performs a 2-port ECal (proceed to Step 5).
Chapter 4
Function
Performs a 1-port calibration on test port 1.
Performs a 1-port calibration on test port 2.
Performs a 1-port calibration on test port 3.
*1
Performs a 1-port calibration on test port 4.
*2
.
.
ECal
. When using a 2-port E5070A/E5071A (Option 213 or 214), pressing
2 Port ECal
ECal (Electronic Calibration)
and confirm that the display turns
Isolation
Calibration
.
ON
103