Operating Concepts
Calibration Routines
Enhanced Reflection Calibration
The enhanced reflection calibration is activated by selecting
ENHANCED RESPONSE
The enhanced reflection calibration effectively removes load match error from the enhanced response
calibration performed on a bilateral device. A bilateral device has an identical forward (S
transmission (S
) response. Most passive devices (such as filters, attenuators, or switches) are bilateral.
12
Some passive devices (circulators, isolators) and most active devices do not have identical forward and
reverse transmission responses and enhanced reflection calibration will not work with these devices.
S
and S
One-Port Calibration
11
22
The S
and S
one-port calibration procedures, activated by pressing the
11
22
softkey within the calibrate menu, provide directivity, source match, and frequency
S22 1-PORT
response vector error-correction for reflection measurements. These procedures provide high accuracy
reflection measurements of one-port devices or properly terminated two-port devices.

Full Two-Port Calibration (ES Models Only)

The full two-port calibration, activated by pressing the
provides directivity, source match, load match, isolation, and frequency response vector error-correction, in
both forward and reverse directions, for transmission and reflection measurements of two-port devices. This
calibration provides the best magnitude and phase measurement accuracy for both transmission and
reflection measurements of two-port devices, and requires an S-parameter test set.
In this type of calibration, both forward and reverse measurements must be made. You have the option of
setting the ratio of the number of forward (or reverse) sweeps versus the number of reverse (or forward)
sweeps. To access this function, press
desired.

TRL*/LRM* Two-Port Calibration

The TRL*/LRM* two-port calibration, activated by pressing the
calibration menu, provides the ability to make calibrations using the TRL or LRM method. For more
information, refer to
"TRL*/LRM* Calibration (ES Models Only)" on page

E-CAL

The E-Cal calibration menu is activated by pressing
The E-Cal (Electronic Calibration) system determines systemic errors of the analyzer through a one-time
connection of an E-Cal module to the network analyzer ports. The random error of connector repeatability is
reduced substantially through a one-time connection when compared to frequent connections and
disconnections of the conventional short/open/load methods.
7- 56
menu.
FULL 2-PORT
Cal MORE
TESTSET SW
E-CAL MENU
ENH. REFL. ON off
S11 1-PORT
softkey within the calibrate menu,
and enter the number of sweeps
TRL*/LRM* 2-PORT
7-66.
in the calibration menu.
under the
) and reverse
21
or
softkey within the