Table 15-5
Corrected System Performance With 3.5 mm Device Connector Type, 85093C
Electronic Calibration Module
Network analyzer: E5070A/E5071A, Calibration module: 85093C (3.5 mm, 50 W) electronic calibration (ECal)
module, Calibration: full 2-port
IF bandwidth = 10 Hz, No averaging applied to data, Environmental temperature = 23°C±5°C with <1°C deviation
from calibration temperature, Isolation calibration not omitted
Description
Directivity
Source Match
Load Match
Reflection Tracking
Transmission Tracking
Transmission Uncertainty (Specification)
Reflection Uncertainty (Specification)
Chapter 15
Specification (dB)
3 MHz to 3 GHz
52
44
47
±0.030
±0.039
Specifications and Supplemental Information
Corrected System Performance
3 GHz to 6 GHz
6 GHz to 8.5 GHz
51
39
44
±0.050
±0.069
47
34
40
±0.070
±0.117