Agilent Technologies ENA Series User Manual page 455

Rf network analyzer
Hide thumbs Also See for ENA Series:
Table of Contents

Advertisement

Key Operation
(Continued)
Displays the segment sweep setup table as well as softkeys for editing the segment table.
Edit Segment Table
Alternates the setup mode for the sweep range between two methods: one
Freq Mode
using the starting and ending frequencies (
the center frequency and a frequency span (
Enables or disables the IF bandwidth display in the segment table.
List IFBW
Enables or disables the power level display in the segment table.
List Power
Enables or disables the segment delay display in the segment table.
List Delay
Enables or disables the segment sweep mode display in the segment table.
List Sweep Mode
Enables or disables the segment sweep time display in the segment table.
List Time
Deletes from the segment table the segment upon which the cursor is
Delete
currently located. If the cursor is not displayed, the segment on the bottom
will be deleted.
Adds a new segment immediately above the segment upon which the cursor
Add
is currently located. If the cursor is not displayed, a new segment will be
added to the bottom of the table.
Displays softkeys for clearing elements of the segment table.
Clear Segment
Table
Deletes all segments in the segment table.
OK
Returns to the softkey display screen one level higher.
Cancel
Exports (saves data in file formats used by other software) the segment
Export to CSV File
table as a CSV (comma-separated value) file.
Imports (loads a file that is written in a format used by other software) a
Import from CSV
CSV (comma-separated value) file to the segment table of
File
E5070A/E5071A.
Returns to the softkey display screen one level higher.
Return
Selects linear frequency (
Segment Display
(1, 2, ..., N; where N refers to the number of points) (
drawing the X-axis of a rectangular display format in a segment sweep.
Returns to the "E5070A/E5071A Menu (Top Menu)" on page 404.
Return
*1.Only with Options 214, 314, and 414.
Appendix D
Function
), and the other using
Start/Stop
).
Center/Span
) or selects the order of measurements
Freq Base
Order Base
Softkey Functions
Sweep Setup Menu
SCPI Command
:SENS{1-9}:SEGM:DATA
:SENS{1-9}:SEGM:DATA
:SENS{1-9}:SEGM:DATA
:SENS{1-9}:SEGM:DATA
:SENS{1-9}:SEGM:DATA
:SENS{1-9}:SEGM:DATA
:SENS{1-9}:SEGM:DATA
:SENS{1-9}:SEGM:DATA
None
:MMEM:STOR:SEGM
:MMEM:LOAD:SEGM
:DISP:WIND{1-9}:X:SPAC LIN
:DISP:WIND{1-9}:X:SPAC OBAS
) for
447

Hide quick links:

Advertisement

Table of Contents